Resuscitation (2005) 67S1, S25—S37 European Resuscitation Council Guidelines for Resuscitation 2005 Section 3. Electrical therapies: Automated external defibrillators, defibrillation, cardioversion and pacing Charles D. Deakin, Jerry P. Nolan Introduction
the rhythm while CPR is in progress is required toprevent unnecessary delays in CPR. Waveform anal-
This section presents guidelines for defibrillation
ysis may also enable the defibrillator to calculate
using both automated external defibrillators (AEDs)
the optimal time at which to give a shock.
and manual defibrillators. All healthcare providersand lay responders can use AEDs as an integral com-ponent of basic life support. Manual defibrillation is
A vital link in the chain of survival
used as part of advanced life support (ALS) therapy. In addition, synchronised cardioversion and pacing
Defibrillation is a key link in the Chain of Survival
are ALS functions of many defibrillators and are also
and is one of the few interventions that have been
shown to improve outcome from VF/VT cardiac
Defibrillation is the passage across the myocard-
arrest. The previous guidelines, published in 2000,
ium of an electrical current of sufficient magnitude
rightly emphasised the importance of early defib-
to depolarise a critical mass of myocardium and
enable restoration of coordinated electrical activ-
The probability of successful defibrillation and
ity. Defibrillation is defined as the termination of
subsequent survival to hospital discharge declines
fibrillation or, more precisely, the absence of ven-
rapidly with timeand the ability to deliver
tricular fibrillation/ventricular tachycardia (VF/VT)
early defibrillation is one of the most important
at 5 s after shock delivery; however, the goal of
factors in determining survival from cardiac
attempted defibrillation is to restore spontaneous
arrest. For every minute that passes following
collapse and defibrillation, mortality increases
Defibrillator technology is advancing rapidly. AED
7%—10% in the absence of bystander CPR.
interaction with the rescuer through voice prompts
systems do not generally have the capability to
is now established, and future technology may
deliver defibrillation through traditional paramedic
enable more specific instructions to be given by
responders within the first few minutes of a call,
voice prompt. The ability of defibrillators to assess
and the alternative use of trained lay responders
0300-9572/$ — see front matter 2005 European Resuscitation Council. All Rights Reserved. Published by Elsevier Ireland Ltd.
to deliver prompt defibrillation using AEDs is now
Automated external defibrillators have been
widespread. EMS systems that have reduced time to
tested extensively against libraries of recorded
defibrillation following cardiac arrest using trained
cardiac rhythms and in many trials in adults
lay responders have reported greatly improved
rhythm analysis. Although AEDs are not designed to
75% if defibrillation is performed within 3 min of
deliver synchronised shocks, all AEDs will recom-
collapse.concept has also been extended to
mend shocks for VT if the rate and R-wave mor-
in-hospital cardiac arrests where staff, other than
doctors, are also being trained to defibrillate usingan AED before arrival of the cardiac arrest team. In-hospital use of AEDs
When bystander CPR is provided, the reduction insurvival rate is more gradual and averages 3%—4%
At the time of the 2005 Consensus Confer-
per minute from collapse to defibrillation;
ence, there were no published randomised trials
comparing in-hospital use of AEDs with manual
defibrillators. Two lower level studies of adults
with in-hospital cardiac arrest from shockable
All healthcare providers with a duty to perform
rhythms showed higher survival-to-hospital dis-
CPR should be trained, equipped, and encouraged
charge rates when defibrillation was provided
to perform defibrillation and CPR. Early defibrilla-
through an AED programme than with manual defib-
tion should be available throughout all hospitals,
rillation alone.manikin study showed that
outpatient medical facilities and public areas of
use of an AED significantly increased the likelihood
mass gathering (see Section 2). Those trained in
of delivering three shocks, but increased the time
AED use should also be trained to deliver at least
to deliver the shocks when compared with manual
external chest compressions before the arrival of
defibrillators.contrast, a study of mock arrests
ALS providers, to optimise the effectiveness of early
in simulated patients showed that use of monitor-
ing leads and fully automated defibrillators reducedtime to defibrillation when compared with manual
Automated external defibrillators
Delayed defibrillation may occur when patients
sustain cardiac arrest in unmonitored hospital beds
Automated external defibrillators are sophisti-
and in outpatient departments. In these areas sev-
cated, reliable computerised devices that use voice
eral minutes may elapse before resuscitation teams
and visual prompts to guide lay rescuers and health-
arrive with a defibrillator and deliver
care professionals to safely attempt defibrillation
Despite limited evidence, AEDs should be consid-
in cardiac arrest victims. Automated defibrillators
ered for the hospital setting as a way to facilitate
have been described as ‘‘. . . the single greatest
early defibrillation (a goal of <3 min from collapse),
advance in the treatment of VF cardiac arrest since
especially in areas where staff have no rhythm
recognition skills or where they use defibrillators
ogy, particularly with respect to battery capacity,
infrequently. An effective system for training and
and software arrhythmia analysis have enabled the
retraining should be in place. Adequate numbers of
mass production of relatively cheap, reliable and
staff should be trained to enable achievement of
the goal of providing the first shock within 3 min of
AEDs by lay or non-healthcare rescuers is covered
collapse anywhere in the hospital. Hospitals should
monitor collapse-to-first-shock intervals and resus-citation outcomes. Automated rhythm analysis
Automated external defibrillators have micropro-cessors that analyse several features of the ECG,
Strategies before defibrillation
including frequency and amplitude. Some AEDs areprogrammed to detect spontaneous movement by
Safe use of oxygen during defibrillation
the patient or others. Developing technology shouldsoon enable AEDs to provide information about
In an oxygen-enriched atmosphere, sparking from
frequency and depth of chest compressions dur-
poorly applied defibrillator paddles can cause a
ing CPR that may improve BLS performance by all
caused in this way, and most have resulted in
European Resuscitation Council Guidelines for Resuscitation 2005
significant burns to the patient. The risk of fire dur-
Shaving the chest
ing attempted defibrillation can be minimised bytaking the following precautions.
Patients with a hairy chest have air trappingbeneath the electrode and poor electrode-to-skin
• Take off any oxygen mask or nasal cannulae and
electrical contact. This causes high impedance,
place them at least 1 m away from the patient’s
reduced defibrillation efficacy, risk of arcing
(sparks) from electrode to skin and electrode
• Leave the ventilation bag connected to the tra-
to electrode and is more likely to cause burns
cheal tube or other airway adjunct. Alterna-
to the patient’s chest. Rapid shaving of the
tively, disconnect any bag-valve device from the
area of intended electrode placement may be
tracheal tube (or other airway adjunct such as
necessary, but do not delay defibrillation if a
the laryngeal mask airway, combitube or laryn-
shaver is not immediately available. Shaving the
geal tube), and remove it at least 1 m from the
chest per se may reduce transthoracic impedance
patient’s chest during defibrillation.
slightly and has been recommended for elective DC
• If the patient is connected to a ventilator, for
example in the operating room or critical careunit, leave the ventilator tubing (breathing cir-
Paddle force
cuit) connected to the tracheal tube unless chestcompressions prevent the ventilator from deliv-
If using paddles, apply them firmly to the chest
ering adequate tidal volumes. In this case, the
wall. This reduces transthoracic impedance by
ventilator is usually substituted for a ventila-
improving electrical contact at the electrode—skin
tion bag, which can itself be left connected or
interface and reducing thoracic volume.The
detached and removed to a distance of at least
defibrillator operator should always press firmly
1 m. If the ventilator tubing is disconnected,
on handheld electrode paddles, the optimal force
ensure it is kept at least 1 m from the patient
being 8 kg in adultsand 5 kg in children aged
or, better still, switch the ventilator off; mod-
1—8 years when using adult paddles8-kg force
ern ventilators generate massive oxygen flows
may be attainable only by the strongest mem-
when disconnected. During normal use, when
bers of the cardiac arrest team, and therefore it
connected to a tracheal tube, oxygen from a ven-
is recommended that these individuals apply the
tilator in the critical care unit will be vented from
paddles during defibrillation. Unlike self-adhesive
the main ventilator housing well away from the
pads, manual paddles have a bare metal plate that
defibrillation zone. Patients in the critical care
requires a conductive material placed between the
unit may be dependent on positive end expiratory
metal and patient’s skin to improve electrical con-
pressure (PEEP) to maintain adequate oxygena-
tact. Use of bare metal paddles alone creates high
tion; during cardioversion, when the spontaneous
transthoracic impedance and is likely to increase
circulation potentially enables blood to remain
the risk of arcing and to worsen cutaneous burns
well oxygenated, it is particularly appropriate to
leave the critically ill patient connected to theventilator during shock delivery. Electrode position
Minimise the risk of sparks during defibrillation. Theoretically, self-adhesive defibrillation pads
No human studies have evaluated the electrode
are less likely to cause sparks than manual pad-
position as a determinant of return of spontaneous
circulation (ROSC) or survival from VF/VT cardiacarrest. Transmyocardial current during defibrilla-
The technique for electrode contact with
tion is likely to be maximal when the electrodes
the chest
are placed so that the area of the heart that is fib-rillating lies directly between them, i.e., ventricles
Optimal defibrillation technique aims to deliver
in VF/VT, atria in atrial fibrillation (AF). Therefore,
current across the fibrillating myocardium in the
the optimal electrode position may not be the same
presence of minimal transthoracic impedance.
for ventricular and atrial arrhythmias.
Transthoracic impedance varies considerably with
More patients are presenting with implantable
medical devices (e.g., permanent pacemaker,
automatic implantable cardioverter defibrillator
place external electrodes (paddles or self-adhesive
(AICD)). MedicAlert bracelets are recommended for
pads) in an optimal position using techniques that
such patients. These devices may be damaged dur-
ing defibrillation if current is discharged through
electrodes placed directly over the device. Place
the electrode away from the device or use an alter-
have shown that anteroposterior electrode place-
native electrode position as described below. On
ment is more effective than the traditional antero-
detecting VF/VT, AICD devices will discharge no
apical position in elective cardioversion of atrial
more than six times. Further discharges will occur
fibrillation. Efficacy of cardioversion may be less
only if a new episode of VF/VT is detected. Rarely,
dependent on electrode position when using bipha-
a faulty device or broken lead may cause repeated
sic impedance-compensated waveforms.Either
firing; in these circumstances, the patient is likely
position is safe and effective for cardioversion of
to be conscious, with the ECG showing a relatively
normal rate. A magnet placed over the AICD willdisable the defibrillation function in these circum-
Respiratory phase
stances. AICD discharge may cause pectoral musclecontraction, but an attendant touching the patient
Transthoracic impedance varies during respiration,
will not receive an electric shock. AICD and pacing
being minimal at end expiration. If possible, defib-
function should always be re-evaluated following
rillation should be attempted at this phase of
external defibrillation, both to check the device
the respiratory cycle. Positive end-expiratory pres-
itself and to check pacing/defibrillation thresholds
sure (PEEP) increases transthoracic impedance and
should be minimised during defibrillation. Auto-
Transdermal drug patches may prevent good
PEEP (gas trapping) may be particularly high in
electrode contact, causing arcing and burns if the
asthmatics and may necessitate higher than usual
electrode is placed directly over the patch during
defibrillation.emove medication patches andwipe the area before applying the electrode. Electrode size
For ventricular arrhythmias, place electrodes
(either pads or paddles) in the conventional
The Association for the Advancement of Medi-
sternal—apical position. The right (sternal) elec-
cal Instrumentation recommends a minimum elec-
trode is placed to the right of the sternum, below
trode size of for individual electrodes and the
the clavicle. The apical paddle is placed in the mid-
sum of the electrode areas should be a mini-
axillary line, approximately level with the V6 ECG
electrode or female breast. This position should
impedance, but excessively large electrodes may
be clear of any breast tissue. It is important that
result in less transmyocardial current flow.
this electrode is placed sufficiently laterally. Other
adult defibrillation, both handheld paddle elec-
trodes and self-adhesive pad electrodes 8—12 cm
in diameter are used and function well. Defib-
each electrode on the lateral chest wall, one
rillation success may be higher with electrodes
on the right and the other on the left side (bi-
of 12-cm diameter compared with those of 8-cm
one electrode in the standard apical position and
Standard AEDs are suitable for use in children
the other on the right or left upper back;
over the age of 8 years. In children between 1 and
one electrode anteriorly, over the left pre-
8 years, use paediatric pads with an attenuator
cordium, and the other electrode posterior to the
to reduce delivered energy, or a paediatric mode,
heart just inferior to the left scapula.
if they are available; if not, use the unmodified
It does not matter which electrode (apex/
machine, taking care to ensure that the adult pads
sternum) is placed in either position.
do not overlap. Use of AEDs is not recommended in
Transthoracic impedance has been shown to be
minimised when the apical electrode is not placedover the female breast.shaped
Coupling agents
apical electrodes have a lower impedance whenplaced longitudinally rather than transversely.
If using manual paddles, gel pads are preferable
The long axis of the apical paddle should therefore
to electrode pastes and gels because the latter
be orientated in a craniocaudal direction.
can spread between the two paddles, creating the
Atrial fibrillation is maintained by functional
potential for a spark. Do not use bare electrodes
re-entry circuits anchored in the left atrium. As
without a coupling material, because this causes
the left atrium is located posteriorly in the tho-
high transthoracic impedance and may increase the
rax, an anteroposterior electrode position may be
severity of any cutaneous burns. Do not use med-
more efficient for external cardioversion of atrial
ical gels or pastes of poor electrical conductivity
European Resuscitation Council Guidelines for Resuscitation 2005
(e.g., ultrasound gel). Electrode pads are preferred
with immediate defibrillation. In contrast, a sin-
to electrode gel because they avoid the risk of
gle randomised study in adults with out-of-hospital
smearing gel between the two paddles and the sub-
VF or VT failed to show improvements in ROSC or
sequent risk of arcing and ineffective defibrillation.
survival following 1.5 min of paramedic animal studies of VF lasting at least 5 min, CPR
Pads versus paddles
before defibrillation improved haemodynamics andsurvival.may not be possible to extrapo-
Self-adhesive defibrillation pads are safe and effec-
late the outcomes achieved by paramedic-provided
tive and are preferable to standard defibrillation
CPR, which includes intubation and delivery of 100%
paddles.Consideration should be given to use
oxygen,those that may be achieved by laypeo-
of self-adhesive pads in peri-arrest situations and
ple providing relative poor-quality CPR with mouth-
in clinical situations where patient access is diffi-
cult. They have a similar transthoracic impedance
It is reasonable for EMS personnel to give a period
of about 2 min of CPR (i.e., about five cycles at
and enable the operator to defibrillate the patient
30:2) before defibrillation in patients with pro-
from a safe distance rather than leaning over the
longed collapse (>5 min). The duration of collapse
patient (as occurs with paddles). When used for ini-
is frequently difficult to estimate accurately, and it
tial monitoring of a rhythm, both pads and paddles
may be simplest if EMS personnel are instructed to
enable quicker delivery of the first shock compared
provide this period of CPR before attempted defib-
with standard ECG electrodes, but pads are quicker
rillation in any cardiac arrest they have not wit-
nessed. Given the relatively weak evidence avail-
When gel pads are used with paddles, the elec-
able, individual EMS directors should determine
trolyte gel becomes polarised and thus is a poor
whether to implement a CPR-before-defibrillation
conductor after defibrillation. This can cause spu-
strategy; inevitably, protocols will vary depending
rious asystole that may persist for 3—4 min when
used to monitor the rhythm; a phenomenon not
Laypeople and first responders using AEDS should
deliver the shock as soon as possible.
a gel pad/paddle combination, confirm a diagnosis
There is no evidence to support or refute CPR
of asystole with independent ECG electrodes rather
before defibrillation for in-hospital cardiac arrest.
We recommend shock delivery as soon as possiblefollowing in-hospital cardiac arrest (see Section 4b
Fibrillation waveform analysis
The importance of early uninterrupted external
It is possible to predict, with varying reliability,
chest compression is emphasised throughout these
the success of defibrillation from the fibrillation
guidelines. In practice, it is often difficult to ascer-
waveform.optimal defibrillation waveforms
tain the exact time of collapse and, in any case,
and the optimal timing of shock delivery can be
CPR should be started as soon as possible. The res-
determined in prospective studies, it should be pos-
cuer providing chest compressions should interrupt
sible to prevent the delivery of unsuccessful high-
chest compressions only for rhythm analysis and
energy shocks and minimise myocardial injury. This
shock delivery, and should be prepared to resume
technology is under active development and inves-
chest compressions as soon as a shock is delivered.
When two rescuers are present, the rescuer operat-ing the AED should apply the electrodes while CPR
CPR versus defibrillation as the initial
is in progress. Interrupt CPR only when it is nec-
treatment
essary to assess the rhythm and deliver a shock. The AED operator should be prepared to deliver a
Although the previous guidelines have recom-
shock as soon as analysis is complete and the shock
mended immediate defibrillation for all shockable
is advised, ensuring all rescuers are not in contact
rhythms, recent evidence has suggested that a
with the victim. The single rescuer should practice
period of CPR before defibrillation may be bene-
coordination of CPR with efficient AED operation.
ficial after prolonged collapse. In clinical studieswhere response times exceeded 4—5 min, a periodof 1.5—3 min of CPR by paramedics or EMS physi-
One-shock versus three-shock sequence
cians before shock delivery improved ROSC, sur-vival to hospital discharge1-year survival
There are no published human or animal studies
for adults with out-of-hospital VF or VT, compared
comparing a single-shock protocol with a three-
stacked-shock protocol for treatment of VF car-diac arrest. Animal studies show that relativelyshort interruptions in external chest compressionto deliver rescue perform rhythmanalysisare associated with post-resuscitationmyocardial dysfunction and reduced survival. Interruptions in external chest compression alsoreduce the chances of converting VF to anotherof CPR performance during out-in-hospitalarrest hasshown that significant interruptions are common,with external chest compressions comprising nomore than 51%total CPR time.
In the context of a three-shock protocol being
recommended in the 2000 guidelines, interruptionsin CPR to enable rhythm analysis by AEDs were
Figure 3.1 Monophasic damped sinusoidal waveform
significant. Delays of up to 37 s between delivery
of shocks and recommencing chest compressionshave been ith first shock efficacy of
of repetitive shocks, which in turn limits myocardial
cardiovert VF successfully is more likely to suggest
After a cautious introduction a decade ago,
the need for a period of CPR rather than a fur-
defibrillators delivering a shock with a biphasic
ther shock. Thus, immediately after giving a single
waveform are now preferred. Monophasic defibril-
shock, and without reassessing the rhythm or feel-
lators are no longer manufactured, although many
ing for a pulse, resume CPR (30 compressions:2 ven-
remain in use. Monophasic defibrillators deliver cur-
tilations) for 2 min before delivering another shock
rent that is unipolar (i.e., one direction of cur-
(if indicated) (see Section 4c). Even if the defibril-
rent flow). There are two main types of monopha-
lation attempt is successful in restoring a perfusing
sic waveform. The commonest waveform is the
rhythm, it is very rare for a pulse to be palpable
monophasic damped sinusoidal (MDS) waveform
immediately after defibrillation, and the delay in
trying to palpate a pulse will further compromise
flow. The monophasic truncated exponential (MTE)
the myocardium if a perfusing rhythm has not been
waveform is electronically terminated before cur-
restored.one study of AEDs in out-of-hospital
VF cardiac arrest, a pulse was detected in only 2.5%
rillators, in contrast, deliver current that flows in
(12/481) of patients with the initial post shock pulse
a positive direction for a specified duration before
check, though a pulse was detected sometime after
reversing and flowing in a negative direction for the
the initial shock sequence (and before a second
remaining milliseconds of the electrical discharge.
There are two main types of biphasic waveform: the
a perfusing rhythm has been restored, giving chest
compressions does not increase the chance of VF
recurring.the presence of post-shock asystole
sic defibrillators compensate for the wide varia-
tions in transthoracic impedance by electronically
This single shock strategy is applicable to both
monophasic and biphasic defibrillators. Waveforms and energy levels
Defibrillation requires the delivery of sufficientelectrical energy to defibrillate a critical mass ofmyocardium, abolish the wavefronts of VF andenable restoration of spontaneous synchronisedelectrical activity in the form of an organisedrhythm. The optimal energy for defibrillation isthat which achieves defibrillation while causing theminimum of myocardial of an
Figure 3.2 Monophasic truncated exponential wave-
appropriate energy level also reduces the number
European Resuscitation Council Guidelines for Resuscitation 2005
The optimal current for defibrillation using amonophasic waveform is in the range of 30—40 A. Indirect evidence from measurements during car-dioversion for atrial fibrillation suggest that the cur-rent during defibrillation using biphasic waveformsis in the range of 15—20 A.Future technologymay enable defibrillators to discharge according totransthoracic current: a strategy that may lead togreater consistency in shock success. Peak currentamplitude, average current and phase duration allneed to be studied to determine optimal values,
Figure 3.3 Biphasic truncated exponential waveform
and manufacturers are encouraged to explore fur-
ther this move from energy-based to current-baseddefibrillation.
adjusting the waveform magnitude and duration. The optimal ratio of first-phase to second-phase
First shock
duration and leading-edge amplitude has not been
First-shock efficacy for long-duration cardiac arrest
established. Whether different waveforms have dif-
using monophasic defibrillation has been reported
fering efficacy for VF of differing durations is also
as 54%—63% for a 200-J monophasic truncated
All manual defibrillators and AEDs that allow
using a 200-J monophasic damped sinusoidal (MDS)
manual override of energy levels should be labelled
to indicate their waveform (monophasic or bipha-
of this waveform, the recommended initial energy
sic) and recommended energy levels for attempted
level for the first shock using a monophasic defib-
defibrillation of VF/VT. First-shock efficacy for
rillator is 360 J. Although higher energy levels risk
long-duration VF/VT is greater with biphasic than
a greater degree of myocardial injury, the bene-
fits of earlier conversion to a perfusing rhythm are
the former is recommended whenever possible.
paramount. Atrioventricular block is more common
Optimal energy levels for both monophasic and
with higher monophasic energy levels, but is gen-
biphasic waveforms are unknown. The recommen-
erally transient and has been shown not to affect
dations for energy levels are based on a consensus
following careful review of the current literature.
studies demonstrated harm caused by attempted
Although energy levels are selected for defibril-
lation, it is the transmyocardial current flow that
There is no evidence that one biphasic wave-
achieves defibrillation. Current correlates well with
form or device is more effective than another. First-
the successful defibrillation and cardioversion.
shock efficacy of the BTE waveform using 150—200 Jhas been reported as 86%—98%.shock efficacy of the RLB waveform using 120 J is upto 85% (data not published in the paper but sup-plied by personnel initialbiphasic shock should be no lower than 120 J forRLB waveforms and 150 J for BTE waveforms. Ide-ally, the initial biphasic shock energy should be atleast 150 J for all waveforms.
Manufacturers should display the effective wave-
form dose range on the face of the biphasic device. If the provider is unaware of the effective doserange of the device, use a dose of 200 J for thefirst shock. This 200 J default energy has been cho-sen because it falls within the reported range ofselected doses that are effective for first and subse-quent biphasic shocks and can be provided by everybiphasic manual defibrillator available today. It isa consensus default dose and not a recommended
Figure 3.4 Rectilinear biphasic waveform (RLB).
ideal dose. If biphasic devices are clearly labelled
and providers are familiar with the devices they use
Blind defibrillation
in clinical care, there will be no need for the default200 J dose. Ongoing research is necessary to firmly
Delivery of shocks without a monitor or an ECG
establish the most appropriate initial settings for
rhythm diagnosis is referred to as ‘‘blind’’ defibril-
both monophasic and biphasic defibrillators.
lation. Blind defibrillation is unnecessary. Handheldpaddles with ‘‘quick-look’’ monitoring capabilities
Second and subsequent shocks
on modern manually operated defibrillators arewidely available. AEDs use reliable and proven deci-
With monophasic defibrillators, if the initial shock
has been unsuccessful at 360 J, second and sub-sequent shocks should all be delivered at 360 J. Spurious asystole and occult ventricular
With biphasic defibrillators there is no evidence to
fibrillation
support either a fixed or escalating energy proto-col. Both strategies are acceptable; however, if the
Rarely, coarse VF can be present in some leads, with
first shock is not successful and the defibrillator is
very small undulations seen in the orthogonal leads,
capable of delivering shocks of higher energy, it
which is called occult VF. A flat line that may resem-
is rational to increase the energy for subsequent
ble asystole is displayed; examine the rhythm in
shocks. If the provider is unaware of the effective
two leads to obtain the correct diagnosis. Of more
dose range of the biphasic device and has used the
importance, one study noted that spurious asystole,
default 200 J dose for the first shock, use either
a flat line produced by technical errors (e.g., no
an equal or higher dose for second or subsequent
power, leads unconnected, gain set to low, incor-
shocks, depending on the capabilities of the device.
rect lead selection, or polarisation of electrolyte
If a shockable rhythm (recurrent ventricular fib-
gel (see above)), was far more frequent than occult
rillation) recurs after successful defibrillation (with
or without ROSC), give the next shock with the
There is no evidence that attempting to defib-
energy level that had previously been successful.
rillate true asystole is beneficial. Studies inchildrenadultsfailed to show bene-fit from defibrillation of asystole. On the contrary,
Other related defibrillation topics
repeated shocks will cause myocardial injury. Defibrillation of children Precordial thump
Cardiac arrest is less common in children. Aetiology
There are no prospective studies that evaluate
is generally related to hypoxia and trauma.
use of precordial (chest) thump. The rationale for
VF is relatively rare compared with adult cardiac
giving a thump is that the mechanical energy of
arrest, occurring in 7%—15% of paediatric and ado-
the thump is converted to electrical energy, which
children include trauma, congenital heart disease,
The electrical threshold of successful defibrillation
long QT interval, drug overdose and hypothermia.
increases rapidly after the onset of the arrhythmia,
Rapid defibrillation of these patients may improve
and the amount of electrical energy generated falls
below this threshold within seconds. A precordial
The optimal energy level, waveform and shock
thump is most likely to be successful in convert-
sequence are unknown but, as with adults, bipha-
ing VT to sinus rhythm. Successful treatment of
sic shocks appear to be at least as effective as,
VF by precordial thump is much less likely: in all
and less harmful than, monophasic shocks.
the reported successful cases, the precordial thump
The upper limit for safe defibrillation is unknown,
but doses in excess of the previously recommended
maximum of 4 J kg−1 (as high as 9 J kg−1) have
less VT was converted to a perfusing rhythm by
defibrillated children effectively without signifi-
a precordial thump, there are occasional reports
of thump causing deterioration in cardiac rhythm,
energy level for manual monophasic defibrillation
such as rate acceleration of VT, conversion of VT
is 4 J kg−1 for the initial shock and for subsequent
shocks. The same energy level is recommended for
Consider giving a single precordial thump when
manual biphasic defibrillation.with adults, if
cardiac arrest is confirmed rapidly after a wit-
a shockable rhythm recurs, use the energy level for
nessed, sudden collapse and a defibrillator is not
defibrillation that had previously been successful.
immediately to hand. These circumstances are
European Resuscitation Council Guidelines for Resuscitation 2005
most likely to occur when the patient is monitored.
randomised study comparing escalating monophasic
Precordial thump should be undertaken immedi-
energy levels to 360 J and biphasic energy levels to
ately after confirmation of cardiac arrest and only
200 J found no difference in efficacy between the
by healthcare professionals trained in the tech-
two waveforms.initial shock of 120—150 J,
nique. Using the ulnar edge of a tightly clenched
escalating if necessary, is a reasonable strategy
fist, a sharp impact is delivered to the lower half of
the sternum from a height of about 20 cm, followedby immediate retraction of the fist, which creates
Atrial flutter and paroxysmal supraventricular tachycardia Cardioversion
require less energy than atrial fibrillation forcardioversion.Give an initial shock of 100-J
If electrical cardioversion is used to convert atrial
monophasic or 70—120 J biphasic waveform. Give
or ventricular tachyarrhythmias, the shock must be
subsequent shocks using stepwise increases in
synchronised to occur with the R wave of the elec-
trocardiogram rather than with the T wave: VF canbe induced if a shock is delivered during the rel-
Ventricular tachycardia
ative refractory portion of the cardiac cycle.Synchronisation can be difficult in VT because of
The energy required for cardioversion of VT
the wide-complex and variable forms of ventricular
depends on the morphological characteristics and
arrhythmia. If synchronisation fails, give unsynchro-
rate of the arrhythmia.entricular tachycardia
nised shocks to the unstable patient in VT to avoid
with a pulse responds well to cardioversion using
prolonged delay in restoring sinus rhythm. Ventric-
initial monophasic energies of 200 J. Use biphasic
ular fibrillation or pulseless VT requires unsynchro-
energy levels of 120—150 J for the initial shock.
nised shocks. Conscious patients must be anaes-
Give stepwise increases if the first shock fails to
thetised or sedated before attempting synchronised
Atrial fibrillation
Consider pacing in patients with symptomatic
Biphasic waveforms are more effective than
bradycardia refractory to anticholinergic drugs or
other second-line therapy (see Section 4f). Immedi-
ate pacing is indicated, especially when the block is
rillator in preference to a monophasic defibrillator.
at or below the His—Purkinje level. If transthoracicpacing is ineffective, consider transvenous pacing. Monophasic waveforms
Whenever a diagnosis of asystole is made, check theECG carefully for the presence of P waves, because
A study of electrical cardioversion for atrial fib-
this may respond to cardiac pacing. Do not attempt
rillation indicated that 360-J MDS shocks were
pacing for asystole; it does not increase short-term
more effective than 100-J or 200-J MDS shocks.
Although a first shock of 360-J reduces overallenergy requirements for cardioversion, 360 J maycause greater myocardial damage than occurs with
References
lower monophasic energy levels, and this must betaken into consideration. Commence cardioversion
1. American Heart Association in collaboration with Inter-
of atrial fibrillation using an initial energy level of
national Liaison Committee on Resuscitation. Guide-lines 2000 for Cardiopulmonary Resuscitation and Emer-
200 J, increasing in a stepwise manner as necessary.
gency Cardiovascular Care, Part 6: Advanced Cardiovas-cular Life Support: Section 2: Defibrillation. Circulation
Biphasic waveforms
2. Larsen MP, Eisenberg MS, Cummins RO, Hallstrom AP.
More data are needed before specific recommen-
Predicting survival from out-of-hospital cardiac arrest: a
dations can be made for optimal biphasic energy
graphic model. Ann Emerg Med 1993;22:1652—8.
3. Valenzuela TD, Roe DJ, Cretin S, Spaite DW, Larsen
levels. First-shock efficacy of a 70-J biphasic wave-
MP. Estimating effectiveness of cardiac arrest interven-
form has been shown to be significantly greater than
tions: a logistic regression survival model. Circulation
that with a 100-monophasic waveform.
4. Waalewijn RA, de Vos R, Tijssen JGP, Koster RW. Survival
ysis in infants and children. Ann Emerg Med 2003;42:185—
models for out-of-hospital cardiopulmonary resuscitation
from the perspectives of the bystander, the first responder,
21. Cecchin F, Jorgenson DB, Berul CI, et al. Is arrhythmia
and the paramedic. Resuscitation 2001;51:113—22.
detection by automatic external defibrillator accurate for
5. Myerburg RJ, Fenster J, Velez M, et al. Impact of
children? Sensitivity and specificity of an automatic exter-
community-wide police car deployment of automated
nal defibrillator algorithm in 696 pediatric arrhythmias.
external defibrillators on survival from out-of-hospital car-
diac arrest. Circulation 2002;106:1058—64.
22. Zafari AM, Zarter SK, Heggen V, et al. A program encour-
6. Capucci A, Aschieri D, Piepoli MF, Bardy GH, Iconomu
aging early defibrillation results in improved in-hospital
E, Arvedi M. Tripling survival from sudden cardiac
resuscitation efficacy. J Am Coll Cardiol 2004;44:846—52.
arrest via early defibrillation without traditional edu-
23. Destro A, Marzaloni M, Sermasi S, Rossi F. Automatic exter-
nal defibrillators in the hospital as well? Resuscitation
7. van Alem AP, Vrenken RH, de Vos R, Tijssen JG, Koster RW.
24. Domanovits H, Meron G, Sterz F, et al. Successful automatic
Use of automated external defibrillator by first responders
external defibrillator operation by people trained only in
in out of hospital cardiac arrest: prospective controlled
basic life support in a simulated cardiac arrest situation.
8. Valenzuela TD, Bjerke HS, Clark LL, et al. Rapid defibrilla-
25. Cusnir H, Tongia R, Sheka KP, et al. In hospital cardiac
tion by nontraditional responders: the Casino Project. Acad
arrest: a role for automatic defibrillation. Resuscitation
9. Swor RA, Jackson RE, Cynar M, et al. Bystander CPR,
26. Kaye W, Mancini ME, Richards N. Organizing and imple-
ventricular fibrillation, and survival in witnessed, unmon-
menting a hospital-wide first-responder automated exter-
itored out-of-hospital cardiac arrest. Ann Emerg Med
nal defibrillation program: strengthening the in-hospital
chain of survival. Resuscitation 1995;30:151—6.
10. Holmberg M, Holmberg S, Herlitz J. Effect of bystander
27. Miller PH. Potential fire hazard in defibrillation. JAMA
cardiopulmonary resuscitation in out-of-hospital cardiac
arrest patients in Sweden. Resuscitation 2000;47:59—70.
28. Hummel IIIrd RS, Ornato JP, Weinberg SM, Clarke AM. Spark-
11. Monsieurs KG, Handley AJ, Bossaert LL. European Resus-
generating properties of electrode gels used during defib-
citation Council Guidelines 2000 for Automated External
rillation. A potential fire hazard. JAMA 1988;260:3021—4.
Defibrillation. A statement from the Basic Life Support and
29. Fires from defibrillation during oxygen administration.
Automated External Defibrillation Working Group (1) and
approved by the Executive Committee of the European
30. Lefever J, Smith A. Risk of fire when using defibrillation in
Resuscitation Council. Resuscitation 2001;48:207—9.
an oxygen enriched atmosphere. Medical Devices Agency
12. Cummins RO, Eisenberg M, Bergner L, Murray JA. Sensitivity
accuracy, and safety of an automatic external defibrillator.
31. Ward ME. Risk of fires when using defibrillators in an oxygen
enriched atmosphere. Resuscitation 1996;31:173.
13. Davis EA, Mosesso Jr VN. Performance of police first
32. Theodorou AA, Gutierrez JA, Berg RA. Fire attributable
responders in utilizing automated external defibrillation
to a defibrillation attempt in a neonate. Pediatrics
on victims of sudden cardiac arrest. Prehosp Emerg Care
33. Kerber RE, Kouba C, Martins J, et al. Advance prediction of
14. White RD, Vukov LF, Bugliosi TF. Early defibrillation by
transthoracic impedance in human defibrillation and car-
police: initial experience with measurement of critical
dioversion: importance of impedance in determining the
time intervals and patient outcome. Ann Emerg Med
success of low-energy shocks. Circulation 1984;70:303—8.
34. Kerber RE, Grayzel J, Hoyt R, Marcus M, Kennedy J.
15. White RD, Hankins DG, Bugliosi TF. Seven years’ experi-
Transthoracic resistance in human defibrillation. Influence
ence with early defibrillation by police and paramedics
of body weight, chest size, serial shocks, paddle size
in an emergency medical services system. Resuscitation
and paddle contact pressure. Circulation 1981;63:676—
16. Wik L, Kramer-Johansen J, Myklebust H, et al. Quality of
35. Sado DM, Deakin CD, Petley GW, Clewlow F. Comparison
cardiopulmonary resuscitation during out-of-hospital car-
of the effects of removal of chest hair with not doing so
diac arrest. JAMA 2005;293:299—304.
before external defibrillation on transthoracic impedance.
17. Abella BS, Alvarado JP, Myklebust H, et al. Quality of
cardiopulmonary resuscitation during in-hospital cardiac
36. Deakin CD, Sado DM, Petley GW, Clewlow F. Differen-
tial contribution of skin impedance and thoracic volume
18. Kerber RE, Becker LB, Bourland JD, et al. Automatic exter-
to transthoracic impedance during external defibrillation.
nal defibrillators for public access defibrillation: recom-
mendations for specifying and reporting arrhythmia analy-
37. Deakin C, Sado D, Petley G, Clewlow F. Determining the
sis algorithm performance, incorporating new waveforms,
optimal paddle force for external defibrillation. Am J Car-
and enhancing safety. A statement for health professionals
from the American Heart Association Task Force on Auto-
38. Deakin C, Bennetts S, Petley G, Clewlow F. What is the opti-
matic External Defibrillation, Subcommittee on AED Safety
mal paddle force for paediatric defibrillation? Resuscitation
and Efficacy. Circulation 1997;95:1677—82.
19. Dickey W, Dalzell GW, Anderson JM, Adgey AA. The accu-
39. Panacek EA, Munger MA, Rutherford WF, Gardner SF. Report
racy of decision-making of a semi-automatic defibrillator
of nitropatch explosions complicating defibrillation. Am J
during cardiac arrest. Eur Heart J 1992;13:608—15.
20. Atkinson E, Mikysa B, Conway JA, et al. Specificity and
40. Wrenn K. The hazards of defibrillation through nitroglycerin
sensitivity of automated external defibrillator rhythm anal-
patches. Ann Emerg Med 1990;19:1327—8.
European Resuscitation Council Guidelines for Resuscitation 2005
41. Pagan-Carlo LA, Spencer KT, Robertson CE, Dengler A, Bir-
and nonparametric classification of ventricular fibrillation
kett C, Kerber RE. Transthoracic defibrillation: importance
in patients with out-of-hospital cardiac arrest. Circulation
of avoiding electrode placement directly on the female
breast. J Am Coll Cardiol 1996;27:449—52.
60. Eftestol T, Wik L, Sunde K, Steen PA. Effects of cardiopul-
42. Deakin CD, Sado DM, Petley GW, Clewlow F. Is the ori-
monary resuscitation on predictors of ventricular fibrilla-
entation of the apical defibrillation paddle of impor-
tion defibrillation success during out-of-hospital cardiac
tance during manual external defibrillation? Resuscitation
arrest. Circulation 2004;110:10—5.
61. Weaver WD, Cobb LA, Dennis D, Ray R, Hallstrom AP, Copass
43. Kirchhof P, Borggrefe M, Breithardt G. Effect of electrode
MK. Amplitude of ventricular fibrillation waveform and out-
position on the outcome of cardioversion. Card Electro-
come after cardiac arrest. Ann Intern Med 1985;102:53—5.
62. Brown CG, Dzwonczyk R. Signal analysis of the human elec-
44. Kirchhof P, Eckardt L, Loh P, et al. Anterior-posterior ver-
trocardiogram during ventricular fibrillation: frequency
sus anterior-lateral electrode positions for external car-
and amplitude parameters as predictors of successful coun-
dioversion of atrial fibrillation: a randomised trial. Lancet
tershock. Ann Emerg Med 1996;27:184—8.
63. Callaham M, Braun O, Valentine W, Clark DM, Zegans C. Pre-
45. Botto GL, Politi A, Bonini W, Broffoni T, Bonatti R. Exter-
hospital cardiac arrest treated by urban first-responders:
nal cardioversion of atrial fibrillation: role of paddle posi-
profile of patient response and prediction of outcome
tion on technical efficacy and energy requirements. Heart
by ventricular fibrillation waveform. Ann Emerg Med
46. Alp NJ, Rahman S, Bell JA, Shahi M. Randomised comparison
64. Strohmenger HU, Lindner KH, Brown CG. Analysis of
of antero-lateral versus antero-posterior paddle positions
the ventricular fibrillation ECG signal amplitude and fre-
for DC cardioversion of persistent atrial fibrillation. Int J
quency parameters as predictors of countershock success
47. Mathew TP, Moore A, McIntyre M, et al. Randomised com-
65. Strohmenger HU, Eftestol T, Sunde K, et al. The predictive
parison of electrode positions for cardioversion of atrial
value of ventricular fibrillation electrocardiogram signal
fibrillation. Heart 1999;81:576—9.
frequency and amplitude variables in patients with out-
48. Walsh SJ, McCarty D, McClelland AJ, et al. Impedance com-
of-hospital cardiac arrest. Anesth Analg 2001;93:1428—33.
pensated biphasic waveforms for transthoracic cardiover-
66. Podbregar M, Kovacic M, Podbregar-Mars A, Brezocnik M.
sion of atrial fibrillation: a multi-centre comparison of
Predicting defibrillation success by ‘genetic’ programming
antero-apical and antero-posterior pad positions. Eur Heart
in patients with out-of-hospital cardiac arrest. Resuscita-
49. Deakin CD, McLaren RM, Petley GW, Clewlow F, Dalrymple-
67. Menegazzi JJ, Callaway CW, Sherman LD, et al. Ventricular
Hay MJ. Effects of positive end-expiratory pressure on
fibrillation scaling exponent can guide timing of defibrilla-
transthoracic impedance—–implications for defibrillation.
tion and other therapies. Circulation 2004;109:926—31.
68. Povoas HP, Weil MH, Tang W, Bisera J, Klouche K, Barbatsis
50. American National Standard: Automatic External Defibril-
A. Predicting the success of defibrillation by electrocardio-
lators and Remote Controlled Defibrillators (DF39). Arling-
graphic analysis. Resuscitation 2002;53:77—82.
ton, Virgina: Association for the Advancement of Medical
69. Noc M, Weil MH, Tang W, Sun S, Pernat A, Bisera J. Electro-
cardiographic prediction of the success of cardiac resusci-
51. Deakin CD, McLaren RM, Petley GW, Clewlow F, Dalrymple-
tation. Crit Care Med 1999;27:708—14.
Hay MJ. A comparison of transthoracic impedance using
70. Strohmenger HU, Lindner KH, Keller A, Lindner IM, Pfen-
standard defibrillation paddles and self-adhesive defibril-
ninger EG. Spectral analysis of ventricular fibrillation and
lation pads. Resuscitation 1998;39:43—6.
closed-chest cardiopulmonary resuscitation. Resuscitation
52. Stults KR, Brown DD, Cooley F, Kerber RE. Self-adhesive
monitor/defibrillation pads improve prehospital defibrilla-
71. Noc M, Weil MH, Gazmuri RJ, Sun S, Biscera J, Tang W.
tion success. Ann Emerg Med 1987;16:872—7.
Ventricular fibrillation voltage as a monitor of the effec-
53. Kerber RE, Martins JB, Kelly KJ, et al. Self-adhesive preap-
tiveness of cardiopulmonary resuscitation. J Lab Clin Med
plied electrode pads for defibrillation and cardioversion. J
72. Lightfoot CB, Nremt P, Callaway CW, et al. Dynamic nature
54. Kerber RE, Martins JB, Ferguson DW, et al. Exper-
of electrocardiographic waveform predicts rescue shock
imental evaluation and initial clinical application of
outcome in porcine ventricular fibrillation. Ann Emerg Med
new self-adhesive defibrillation electrodes. Int J Cardiol
73. Marn-Pernat A, Weil MH, Tang W, Pernat A, Bisera J. Opti-
55. Perkins GD, Roberts C, Gao F. Delays in defibrillation: influ-
mizing timing of ventricular defibrillation. Crit Care Med
ence of different monitoring techniques. Br J Anaesth
74. Hamprecht FA, Achleitner U, Krismer AC, et al. Fibrilla-
56. Bradbury N, Hyde D, Nolan J. Reliability of ECG monitor-
tion power, an alternative method of ECG spectral analysis
ing with a gel pad/paddle combination after defibrillation.
for prediction of countershock success in a porcine model
of ventricular fibrillation. Resuscitation 2001;50:287—
57. Chamberlain D. Gel pads should not be used for monitoring
ECG after defibrillation. Resuscitation 2000;43:159—60.
75. Amann A, Achleitner U, Antretter H, et al. Analysing ven-
58. Callaway CW, Sherman LD, Mosesso Jr VN, Dietrich TJ, Holt
tricular fibrillation ECG-signals and predicting defibrillation
E, Clarkson MC. Scaling exponent predicts defibrillation
success during cardiopulmonary resuscitation employing
success for out-of-hospital ventricular fibrillation cardiac
N(alpha)-histograms. Resuscitation 2001;50:77—85.
arrest. Circulation 2001;103:1656—61.
76. Brown CG, Griffith RF, Van Ligten P, et al. Median
59. Eftestol T, Sunde K, Aase SO, Husoy JH, Steen PA. Predict-
frequency—–a new parameter for predicting defibrillation
ing outcome of defibrillation by spectral characterization
success rate. Ann Emerg Med 1991;20:787—9.
77. Amann A, Rheinberger K, Achleitner U, et al. The predic-
93. Rea TD, Shah S, Kudenchuk PJ, Copass MK, Cobb LA. Auto-
tion of defibrillation outcome using a new combination of
mated external defibrillators: to what extent does the algo-
mean frequency and amplitude in porcine models of car-
rithm delay CPR? Ann Emerg Med 2005;46:132—41.
diac arrest. Anesth Analg 2002;95:716—22.
94. Hess EP, White RD. Ventricular fibrillation is not pro-
78. Cobb LA, Fahrenbruch CE, Walsh TR, et al. Influence
voked by chest compression during post-shock organized
of cardiopulmonary resuscitation prior to defibrillation in
rhythms in out-of-hospital cardiac arrest. Resuscitation
patients with out-of-hospital ventricular fibrillation. JAMA
95. Joglar JA, Kessler DJ, Welch PJ, et al. Effects of repeated
79. Wik L, Hansen TB, Fylling F, et al. Delaying defibrillation to
electrical defibrillations on cardiac troponin I levels. Am J
give basic cardiopulmonary resuscitation to patients with
out-of-hospital ventricular fibrillation: a randomized trial.
96. van Alem AP, Chapman FW, Lank P, Hart AA, Koster
RW. A prospective, randomised and blinded compari-
80. Jacobs IG, Finn JC, Oxer HF, Jelinek GA. CPR before defibril-
son of first shock success of monophasic and biphasic
lation in out-of-hospital cardiac arrest: a randomized trial.
waveforms in out-of-hospital cardiac arrest. Resuscitation
Emerg Med Australas 2005;17:39—45.
81. Berg RA, Hilwig RW, Kern KB, Ewy GA. Precountershock
97. Carpenter J, Rea TD, Murray JA, Kudenchuk PJ, Eisenberg
cardiopulmonary resuscitation improves ventricular fibril-
MS. Defibrillation waveform and post-shock rhythm in out-
lation median frequency and myocardial readiness for suc-
of-hospital ventricular fibrillation cardiac arrest. Resusci-
cessful defibrillation from prolonged ventricular fibrilla-
tion: a randomized, controlled swine study. Ann Emerg Med
98. Morrison LJ, Dorian P, Long J, et al. Out-of-hospital car-
diac arrest rectilinear biphasic to monophasic damped sine
82. Berg RA, Hilwig RW, Ewy GA, Kern KB. Precountershock
defibrillation waveforms with advanced life support inter-
cardiopulmonary resuscitation improves initial response
vention trial (ORBIT). Resuscitation 2005;66:149—57.
to defibrillation from prolonged ventricular fibrillation:
99. Kerber RE, Martins JB, Kienzle MG, et al. Energy, current,
a randomized, controlled swine study. Crit Care Med
and success in defibrillation and cardioversion: clinical
studies using an automated impedance-based method of
83. Kolarova J, Ayoub IM, Yi Z, Gazmuri RJ. Optimal timing for
energy adjustment. Circulation 1988;77:1038—46.
electrical defibrillation after prolonged untreated ventric-
100. Koster RW, Dorian P, Chapman FW, Schmitt PW, O’Grady
ular fibrillation. Crit Care Med 2003;31:2022—8.
SG, Walker RG. A randomized trial comparing monophasic
84. Berg RA, Sanders AB, Kern KB, et al. Adverse hemo-
and biphasic waveform shocks for external cardioversion of
dynamic effects of interrupting chest compressions for
atrial fibrillation. Am Heart J 2004;147:e20.
rescue breathing during cardiopulmonary resuscitation
101. Martens PR, Russell JK, Wolcke B, et al. Optimal Response
for ventricular fibrillation cardiac arrest. Circulation
to Cardiac Arrest study: defibrillation waveform effects.
85. Kern KB, Hilwig RW, Berg RA, Sanders AB, Ewy GA.
102. Weaver WD, Cobb LA, Copass MK, Hallstrom AP. Ventricu-
Importance of continuous chest compressions during
lar defibrillation: a comparative trial using 175-J and 320-J
cardiopulmonary resuscitation: improved outcome dur-
shocks. N Engl J Med 1982;307:1101—6.
ing a simulated single lay-rescuer scenario. Circulation
103. Tang W, Weil MH, Sun S, et al. The effects of
biphasic and conventional monophasic defibrillation on
86. Yu T, Weil MH, Tang W, et al. Adverse outcomes of inter-
postresuscitation myocardial function. J Am Coll Cardiol
rupted precordial compression during automated defibril-
lation. Circulation 2002;106:368—72.
104. Gliner BE, Jorgenson DB, Poole JE, et al. Treatment of out-
87. Eftestol T, Sunde K, Steen PA. Effects of interrupting pre-
of-hospital cardiac arrest with a low-energy impedance-
cordial compressions on the calculated probability of defib-
compensating biphasic waveform automatic external defib-
rillation success during out-of-hospital cardiac arrest. Cir-
rillator. The LIFE Investigators. Biomed Instrum Technol
88. Valenzuela TD, Kern KB, Clark LL, et al. Interruptions
105. White RD, Blackwell TH, Russell JK, Snyder DE, Jorgenson
of chest compressions during emergency medical systems
DB. Transthoracic impedance does not affect defibrillation,
resuscitation. Circulation 2005;112:1259—65.
resuscitation or survival in patients with out-of-hospital
89. van Alem AP, Sanou BT, Koster RW. Interruption of car-
cardiac arrest treated with a non-escalating biphasic wave-
diopulmonary resuscitation with the use of the automated
form defibrillator. Resuscitation 2005;64:63—9.
external defibrillator in out-of-hospital cardiac arrest. Ann
106. Kuisma M, Suominen P, Korpela R. Paediatric out-of-hospital
cardiac arrests: epidemiology and outcome. Resuscitation
90. Bain AC, Swerdlow CD, Love CJ, et al. Multicenter study
of principles-based waveforms for external defibrillation.
107. Sirbaugh PE, Pepe PE, Shook JE, et al. A prospective,
population-based study of the demographics, epidemi-
91. Poole JE, White RD, Kanz KG, et al. Low-energy impedance-
ology, management, and outcome of out-of-hospital
compensating biphasic waveforms terminate ventricular
fibrillation at high rates in victims of out-of-hospital car-
diac arrest. LIFE Investigators. J Cardiovasc Electrophysiol
108. Hickey RW, Cohen DM, Strausbaugh S, Dietrich AM. Pedi-
atric patients requiring CPR in the prehospital setting. Ann
92. Schneider T, Martens PR, Paschen H, et al. Multicenter,
randomized, controlled trial of 150-J biphasic shocks com-
109. Appleton GO, Cummins RO, Larson MP, Graves JR. CPR and
pared with 200- to 360-J monophasic shocks in the resusci-
the single rescuer: at what age should you ‘‘call first’’
tation of out-of-hospital cardiac arrest victims. Optimized
rather than ‘‘call fast’’? Ann Emerg Med 1995;25:492—4.
Response to Cardiac Arrest (ORCA) Investigators. Circula-
110. Ronco R, King W, Donley DK, Tilden SJ. Outcome and cost
at a children’s hospital following resuscitation for out-of-
European Resuscitation Council Guidelines for Resuscitation 2005
hospital cardiopulmonary arrest. Arch Pediatr Adolesc Med
129. Gertsch M, Hottinger S, Hess T. Serial chest thumps for the
treatment of ventricular tachycardia in patients with coro-
111. Losek JD, Hennes H, Glaeser P, Hendley G, Nelson DB.
nary artery disease. Clin Cardiol 1992;15:181—8.
Prehospital care of the pulseless, nonbreathing pediatric
130. Krijne R. Rate acceleration of ventricular tachycardia after
patient. Am J Emerg Med 1987;5:370—4.
a precordial chest thump. Am J Cardiol 1984;53:964—5.
112. Mogayzel C, Quan L, Graves JR, Tiedeman D, Fahrenbruch
131. Sclarovsky S, Kracoff OH, Agmon J. Acceleration of ven-
C, Herndon P. Out-of-hospital ventricular fibrillation in chil-
tricular tachycardia induced by a chest thump. Chest
dren and adolescents: causes and outcomes. Ann Emerg
132. Yakaitis RW, Redding JS. Precordial thumping during cardiac
113. Safranek DJ, Eisenberg MS, Larsen MP. The epidemiol-
resuscitation. Crit Care Med 1973;1:22—6.
ogy of cardiac arrest in young adults. Ann Emerg Med
133. Lown B. Electrical reversion of cardiac arrhythmias. Br
114. Berg RA, Chapman FW, Berg MD, et al. Attenuated adult
134. Mittal S, Ayati S, Stein KM, et al. Transthoracic cardiover-
biphasic shocks compared with weight-based monophasic
sion of atrial fibrillation: comparison of rectilinear biphasic
shocks in a swine model of prolonged pediatric ventricular
versus damped sine wave monophasic shocks. Circulation
fibrillation. Resuscitation 2004;61:189—97.
115. Tang W, Weil MH, Jorgenson D, et al. Fixed-energy bipha-
135. Page RL, Kerber RE, Russell JK, et al. Biphasic versus
sic waveform defibrillation in a pediatric model of cardiac
monophasic shock waveform for conversion of atrial fibril-
arrest and resuscitation. Crit Care Med 2002;30:2736—41.
lation: the results of an international randomized, double-
116. Clark CB, Zhang Y, Davies LR, Karlsson G, Kerber RE. Pedi-
blind multicenter trial. J Am Coll Cardiol 2002;39:1956—63.
atric transthoracic defibrillation: biphasic versus monopha-
136. Joglar JA, Hamdan MH, Ramaswamy K, et al. Initial energy
sic waveforms in an experimental model. Resuscitation
for elective external cardioversion of persistent atrial fib-
rillation. Am J Cardiol 2000;86:348—50.
117. Gurnett CA, Atkins DL. Successful use of a biphasic wave-
137. Alatawi F, Gurevitz O, White R. Prospective, randomized
form automated external defibrillator in a high-risk child.
comparison of two biphasic waveforms for the efficacy and
safety of transthoracic biphasic cardioversion of atrial fib-
118. Atkins DL, Jorgenson DB. Attenuated pediatric electrode
rillation. Heart Rhythm 2005;2:382—7.
pads for automated external defibrillator use in children.
138. Pinski SL, Sgarbossa EB, Ching E, Trohman RG. A comparison
of 50-J versus 100-J shocks for direct-current cardioversion
119. Gutgesell HP, Tacker WA, Geddes LA, Davis S, Lie JT, McNa-
of atrial flutter. Am Heart J 1999;137:439—42.
mara DG. Energy dose for ventricular defibrillation of chil-
139. Kerber RE, Kienzle MG, Olshansky B, et al. Ventricular
tachycardia rate and morphology determine energy and
120. Cummins RO, Austin Jr D. The frequency of ‘occult’ ventric-
current requirements for transthoracic cardioversion. Cir-
ular fibrillation masquerading as a flat line in prehospital
cardiac arrest. Ann Emerg Med 1988;17:813—7.
140. Hedges JR, Syverud SA, Dalsey WC, Feero S, Easter R, Shultz
121. Losek JD, Hennes H, Glaeser PW, Smith DS, Hendley G. Pre-
B. Prehospital trial of emergency transcutaneous cardiac
hospital countershock treatment of pediatric asystole. Am
pacing. Circulation 1987;76:1337—43.
141. Barthell E, Troiano P, Olson D, Stueven HA, Hendley G. Pre-
122. Martin DR, Gavin T, Bianco J, et al. Initial countershock in
hospital external cardiac pacing: a prospective, controlled
the treatment of asystole. Resuscitation 1993;26:63—8.
clinical trial. Ann Emerg Med 1988;17:1221—6.
123. Kohl P, King AM, Boulin C. Antiarrhythmic effects of acute
142. Cummins RO, Graves JR, Larsen MP, et al. Out-of-hospital
mechanical stiumulation. In: Kohl P, Sachs F, Franz MR, edi-
transcutaneous pacing by emergency medical technicians
tors. Cardiac mechano-electric feedback and arrhythmias:
in patients with asystolic cardiac arrest. N Engl J Med
form pipette to patient. Philadelphia: Elsevier Saunders;
143. Ornato JP, Peberdy MA. The mystery of bradyasystole
124. Befeler B. Mechanical stimulation of the heart; its thera-
during cardiac arrest. Ann Emerg Med 1996;27:576—
peutic value in tachyarrhythmias. Chest 1978;73:832—8.
144. Niemann JT, Adomian GE, Garner D, Rosborough JP. Endo-
Siegert K. Terminierung von Kammertachykardien durch
cardial and transcutaneous cardiac pacing, calcium chlo-
ride, and epinephrine in postcountershock asystole and
(‘‘Termination of Ventricular Tachycardias by Mechan-
bradycardias. Crit Care Med 1985;13:699—704.
ical Cardiac Pacing by Means of Precordial Thumps’’).
145. Quan L, Graves JR, Kinder DR, Horan S, Cummins RO.
Transcutaneous cardiac pacing in the treatment of out-
126. Caldwell G, Millar G, Quinn E. Simple mechanical meth-
of-hospital pediatric cardiac arrests. Ann Emerg Med
ods for cardioversion: Defence of the precordial thump and
cough version. Br Med J 1985;291:627—30.
146. Dalsey WC, Syverud SA, Hedges JR. Emergency department
127. Morgera T, Baldi N, Chersevani D, Medugno G, Camerini F.
use of transcutaneous pacing for cardiac arrests. Crit Care
Chest thump and ventricular tachycardia. Pacing Clin Elec-
147. Knowlton AA, Falk RH. External cardiac pacing during in-
128. Rahner E, Zeh E. Die Regularisierung von Kammertachykar-
hospital cardiac arrest. Am J Cardiol 1986;57:1295—8.
akordialen Faustschlag. (‘‘The Regulariza-
148. Ornato JP, Carveth WL, Windle JR. Pacemaker insertion for
tion of Ventricular Tachycardias by Precordial Thumping’’).
prehospital bradyasystolic cardiac arrest. Ann Emerg Med
Copyright, 1996, by the Massachusetts Medical Society EFFECT OF AN ENTERIC-COATED FISH-OIL PREPARATION ON RELAPSES IN CROHN’S DISEASE ANDREA BELLUZZI, M.D., CORRADO BRIGNOLA, M.D., MASSIMO CAMPIERI, M.D., ANGELO PERA, M.D., STEFANO BOSCHI, M.S., AND MARIO MIGLIOLI, M.D. Abstract the 39 patients in the placebo group, 27 (69 percent) hadmay have periods o
EffEcts of mEphEnoxalonE and tizanidinE for acutE painful spasm of paravErtEbral musclEs: a randomizEd doublE-blind control study Sheng-Mou Hou, Shu-Hua Yang2,3, Jyh-Horng Wang2 objectives: skeletal muscle relaxants are frequently used to treat low back pain with spasm of paravertebral muscles. however, there is little comparative efficacy and safety data of different muscle re